oxygen sensor for mysign o oxygen analyzer
Click To Enlarge

Oxygen Sensor for MySign O Oxygen Analyzer

Specifications:

Measurement Range: 0-100 % oxygen
Output in ambient air: 11 to 13mV
Electrical Interface: Stereo phone jack (3,5mm)
Accuracy and Repeatability: < 1 % vol. O2 when calibrated at 100 %
Oxygen
Linearity error: < 3 % relative
Response time: < 12sec. to 90 % of final value
Zero Offset Voltage: < 200 µV in 100 % nitrogen
applied after 5 min
Cross Interference: Meets EN ISO 21647 requirements
Influence of Humidity: - 0.03 % rel. per % RH at 25°C
Influence of Pressure: proportional to change in oxygen partial
pressure
Influence of Mechanical Shock: < 1% relative after a fall from 1m
Operating Temperature: 0 to 50°C
Temperature Compensation: built-in NTC compensation
Effect of Temperature
Compensation (steady state):
between +25°C and +40°C: 3 % relative error
between 0 °C and +50 °C: 8 % relative error
Operating Humidity: 0-99 % RH non-condensing
Long Term Output Drift: < 1 % vol oxygen per month
typically < - 15 % relative over lifetime
Storage Temperature: -20 to +50 °C
Recommended Storage: +5 to +15 °C
Recommended Load: ≥ 10 kOhms
Warm-Up Time: < 30 minutes, after replacement of sensor
Nominal Sensor Lifetime: ≥ 1.000.000 % vol oxygen hours
Weight: approximately 28 grams
Sensor coding: Built-in data memory chip
Part No.: 01-00-0114

All specifications are applicable at standard conditions: 1013 hPa, 25°C dry ambient air

Use the advantages:
• Meets EN ISO 21647
• Designed and
manufactured according to
EN ISO 9001 : 2000 and EN
ISO 13485 : 2003
• Sensor coding integrated
• Accurate and reliable
response
• Resistant to N2O
• Excellent signal stability
• High product quality

MADE IN GERMANY

  • Item #: B226-220
  • Condition: New

Oxygen Sensor for MySign O Oxygen Analyzer

Price: $125.00
* Marked fields are required.
Qty: *
Reviews (0) Write a Review
No Reviews. Write a Review